What it is
Some surgical implants need a hole down the middle. A guide wire passes through. The implant follows the wire into the exact position.
Cannulated titanium bar for surgical implants is hollow bar made for that purpose. Not solid. Not tube. Bar with a precision hole. Thick enough wall for strength. Hollow enough for a guide wire.
Cannulated screws. Intramedullary nails. Hip fracture implants. Dental implants. Biopsy needles. Any implant that goes over a guide wire.
Why cannulated
Without cannulation, the surgeon has to guess. With cannulation, the guide wire goes first. X-ray confirms position. Then the implant slides over the wire. Perfect placement every time.
For cannulated titanium bar for surgical implants, the inner hole is what makes minimally invasive surgery possible. Smaller incision. Less radiation. Better outcome.
Which grade for which implant
| Grade | Material | Tensile | Best for |
| Grade 23 ELI | Ti-6Al-4V | 860 MPa | Load-bearing implants. Screws. Nails. |
| Grade 4 CP | Pure titanium | 550 MPa | Non-load-bearing. Biopsy needles. Instruments. |
Grade 23 ELI is the standard for cannulated screws and IM nails. High strength allows thin wall. Small diameter. Load bearing.
Grade 4 CP is used for biopsy needles and surgical instruments. No strength requirement. Pure titanium is sufficient.
For cannulated titanium bar for surgical implants, most load-bearing applications use Grade 23 ELI.
Material specs (Grade 23 ELI)
| Property | Value |
| Standard | ASTM F136 |
| Tensile (min) | 860 MPa |
| Yield (min) | 795 MPa |
| Elongation | ≥ 10% |
| Density | 4.43 g/cm³ |
Chemical limits: Oxygen ≤0.13%, Iron ≤0.25%. No nickel. No aluminum concerns. Medical grade.
Common sizes for surgical implants
|
Outer (mm) |
Inner (mm) |
Wall (mm) | Implant type |
| 1.5 | 0.8 | 0.35 | Micro biopsy needle |
| 2.0 | 1.0 | 0.5 | Small cannulated screw |
| 2.5 | 1.2 | 0.65 | Hand screw |
| 3.0 | 1.5 | 0.75 | Dental guide wire implant |
| 3.5 | 1.8 | 0.85 | Wrist screw |
| 4.0 | 2.0 | 1.0 | Ankle screw |
| 5.0 | 2.5 | 1.25 | Small hip screw |
| 6.0 | 3.0 | 1.5 | Standard hip screw |
| 8.0 | 4.0 | 2.0 | Tibial IM nail |
| 10.0 | 5.0 | 2.5 | Small femoral nail |
| 12.0 | 6.0 | 3.0 | Standard femoral nail |
| 15.0 | 8.0 | 3.5 | Large femoral nail |
| 20.0 | 14.0 | 3.0 | Custom prosthesis |
Wall thickness from 0.35mm for tiny needles to 3.5mm for large nails.
Where it goes
Hip fracture screws. Femoral neck fracture. Guide wire first. Screw follows. Patient can bear weight soon.
IM nails. Long bone fractures. Femur. Tibia. Humerus. Cannulated nail slides over wire. Locking screws at both ends.
Cannulated screws for wrist and ankle. Smaller diameters. 3.5mm to 4.0mm outer. Hand surgery. Foot surgery.
Dental implants. Hollow fixtures for bone ingrowth. Guide wire placement. Precise angle control.
Biopsy needles. Outer cannula. Inner stylet. Tissue sample through the center. Thin wall. Sharp tip.
For cannulated titanium bar for surgical implants, each implant type has different diameter and wall thickness requirements.
What makes cannulated bar different from tube
Tube is thin wall. Made for fluid flow. Not strong enough for load-bearing implants.
Cannulated bar has thicker wall. Made for strength with a hole through the center.
| Feature | Cannulated bar | Medical tube |
| Wall thickness | 0.5mm – 3.5mm | 0.1mm – 1.0mm |
| Strength | High (load-bearing) | Low |
| Primary use | Implants over guide wire | Catheters, fluid flow |
| Concentricity | Tight | Moderate |
For cannulated titanium bar for surgical implants, the wall must be thick enough to hold body weight.
Critical tolerances
| Parameter | Requirement |
| Concentricity | ≤ 0.05mm |
| Inner diameter | ±0.025mm |
| Outer diameter | ±0.025mm |
| Straightness | 0.1mm per 300mm |
| Surface finish (Ra) | ≤ 0.2μm |
These tolerances ensure the implant slides smoothly over the guide wire. No binding. No wobble.
Guide wire fit
The inner diameter must match the guide wire.
| Guide wire | Inner diameter | Clearance |
| 0.8mm – 1.0mm | +0.1mm | Sliding fit |
| 1.5mm – 2.0mm | +0.15mm | Smooth glide |
| 2.5mm – 4.0mm | +0.2mm | Easy passage |
Too tight: implant sticks. Too loose: implant wobbles.
Surface finish inside bore
The inner bore contacts the guide wire. Rough bore catches. Smooth bore glides.
Our process: Drill, ream, hone, then flow polish. Ra ≤ 0.2μm inside and out.
For cannulated titanium bar for surgical implants, the inside is as smooth as the outside.
How cannulated bar is made
Start with solid medical grade rod. Deep hole drilling. The drill bit must stay centered. High pressure coolant pushes chips out.
Multi-step process. Rough drill. Then ream. Then hone. Each step improves concentricity and surface finish.
Centerless grinding. Outer diameter ground to tolerance. Uses the inner bore as reference.
Straightening. Slow pressure. No damage to inner bore.
Final inspection. Camera down every bar. Check for burrs. Check for steps. Check for smoothness.
Manufacturing your implant
Cannulated bar arrives at your facility. Cut to implant length. Machine screw threads. Drill locking holes. Add surface treatment. Sterilize.
For cannulated titanium bar for surgical implants, the bar is ready for your CNC. Consistent diameter. No surprises.
Traceability
Cannulated implants are medical devices. Class II or III. Full traceability required.
Every cannulated titanium bar for surgical implants comes with:
Heat number. Mill test report. ASTM F136 or F67 cert. Chemical analysis. Mechanical test results. Concentricity report. Inner diameter cert.
EN 10204 3.1 standard. 3.2 with third-party witness available.
FAQ
Q: Grade 23 ELI or Grade 4 for cannulated hip screws?
A: Grade 23 ELI. Hip screws take body weight. Need high strength. Grade 23 ELI gives 860 MPa. Grade 4 is 550 MPa.
Q: What wall thickness for a 6.0mm hip screw?
A: 1.5mm. Outer 6.0mm, inner 3.0mm gives 1.5mm wall. Standard for hip screws.
Q: Minimum inner diameter?
A: 0.8mm possible for small guide wires. Larger is more reliable. Ask us.
Q: Can you hold concentricity for long bars?
A: Yes. Up to 2000mm length. Concentricity ≤0.05mm.
Q: Lead time?
A: Stock sizes 4-5 weeks. Custom sizes 6-8 weeks.
Q: Sample bar?
A: Yes. 200mm sample. Pay material and shipping. Credit back on first production order over $20,000.
The bottom line
Cannulated implants go over guide wires. Precise placement. Minimally invasive.
Cannulated titanium bar for surgical implants gives you the hollow starting stock.
Grade 23 ELI for load-bearing screws and nails. Grade 4 CP for non-load-bearing.
Concentric hole. Smooth bore. Strong wall.
For hip screws. For IM nails. For dental implants. For biopsy needles.
Hollow in the middle. Strong on the outside. Ready for your machining.
Contact
Need a quote on cannulated titanium bar for surgical implants? Tell us outer diameter, inner diameter, length, quantity, and grade. We'll reply within 24 hours.
Email: kristy@mt-titanium.com
WhatsApp: +86-17391509251
Hot Tags: cannulated titanium bar for surgical implants, China cannulated titanium bar for surgical implants manufacturers, suppliers, factory, Cannulated Titanium Bar for Surgical Implants, High Strength Cannulated Titanium Bar for Trauma Surgery, Hollow Titanium Bar for Dental Implant Fixtures ASTM F67, Precision Cannulated Titanium Bar, Ti‑6Al‑4V Cannulated Bar for Joint Replacement Components, titanium cannulated bar