ASTM F67 Pure Titanium Rod Gr1 Gr2 Gr3 Gr4 Surgical Implant

Send Inquiry
ASTM F67 Pure Titanium Rod Gr1 Gr2 Gr3 Gr4 Surgical Implant
Details
Material: Grade 1 / Grade 2 / Grade 3 / Grade 4 (UNS R50250 / R50400 / R50550 / R50700)
Standard: ASTM F67 / ISO 5832-2
Diameter: 3.0mm – 100mm
Length: 500mm – 3000mm
Surface: Ground, polished, or turned
Condition: Annealed
Category
Medical Titanium Rod
Share to
Description

ASTM F67 Description

 

ASTM F67 isn't just a spec number. It's the line between material that can go into a human body and material that can't. Industrial titanium is cheap and available. Implant-grade titanium is controlled, tested, and traceable. The FDA has recognized ASTM F67 as a consensus standard for surgical implant applications since the late 1990s, and the current revision F67-24 remains on the recognized list.

The spec covers four grades of unalloyed titanium - Grade 1 through Grade 4 - in strip, sheet, plate, bar, billet, forging, and wire form. Each grade has a different balance of strength and ductility. Grade 1 bends the easiest. Grade 4 carries the most load. Picking the wrong one isn't a cost issue. It's a design failure.

 

Product Specifications

 

Grade

UNS

Tensile (min)

Yield (min)

Elongation

Reduction of Area

Grade 1

R50250

240 MPa

170 MPa

24%

30%

Grade 2

R50400

345 MPa

275 MPa

20%

30%

Grade 3

R50550

450 MPa

380 MPa

18%

30%

Grade 4

R50700

550 MPa

483 MPa

15%

25%

 

Chemical Composition (ASTM F67, max limits)

 

Element

Gr1

Gr2

Gr3

Gr4

Oxygen

0.18

0.25

0.35

0.40

Iron

0.20

0.30

0.30

0.50

Carbon

0.08

0.08

0.08

0.08

Nitrogen

0.03

0.03

0.05

0.05

Hydrogen

0.015

0.015

0.015

0.015

Titanium

Balance

Balance

Balance

Balance

 

The oxygen numbers tell the story. Grade 1 at 0.18% oxygen is soft and formable. Grade 4 at 0.40% oxygen is nearly double the strength but half the ductility. That trade-off is why Grade 4 is used for dental fixtures that need osseointegration and load capacity, while Grade 1 is used for cranial mesh that surgeons shape by hand in the operating room.

 

Production Route - From Sponge to Surgical Rod

 

Implant-grade titanium doesn't start as metal. It starts as titanium sponge - porous, grey, and chemically raw. That sponge goes through a chain of steps that turns it into a certified surgical implant rod. Every step matters. Any shortcut shows up in the final microstructure.

Titanium sponge is the base. It's produced from rutile or ilmenite ore through the Kroll process. The sponge is pressed into a consumable electrode - often called a pizza electrode because of its disc shape - and welded together into a column. That electrode goes into a vacuum arc remelting furnace. The arc melts the electrode. Molten titanium drips into a water-cooled copper crucible. Impurities either vaporize or float out. The ingot that forms is the first solid titanium in the chain.

That ingot gets forged. Forging breaks up the cast structure. It closes porosity. It refines grain size. The result is a rod billet - a semi-finished bar ready for further processing. The billet then goes to hot rolling. Temperature is controlled. Reduction ratios are controlled. The goal is uniform deformation through the cross-section. After rolling, the rod is pickled. Acid bath removes the oxide scale that formed during hot working. Pickling also exposes surface defects so they can be caught before machining.

Machining comes next. The rod is turned or centerless ground to the specified diameter. Tolerance tightens. Surface finish improves. Straightness is corrected. Then polishing. Polished rod has a smooth surface for tissue contact applications. It also removes any remaining surface imperfections that could act as fatigue crack initiation sites.

The final product is titanium rod - certified, tested, and traceable. Every step from sponge to finished rod is documented. The heat number stays with the material through every process. That's how implant-grade traceability works.

Process of the titanium rod

Step

What Happens

Why It Matters

Titanium sponge

Kroll process from ore

Base material. Chemistry starts here.

Pizza electrode

Sponge pressed and welded

Consumable electrode for melting.

Smelting

Vacuum arc remelting

Removes impurities. Homogenizes chemistry.

Forging

Hot deformation

Breaks cast structure. Refines grain.

Rod billet

Semi-finished bar

Intermediate product. Ready for rolling.

Hot rolling

Controlled deformation

Shapes and reduces diameter.

Pickling

Acid bath

Removes oxide scale. Reveals surface defects.

Machining

Turning or centerless grinding

Achieves diameter tolerance.

Polishing

Surface finishing

Smooth surface. Removes crack initiation sites.

Titanium Rod

Final product

Certified. Tested. Traceable.

 

 

Inspection

 

ASTM F67 requires more than a basic mill certificate. Each lot must be tested for chemical composition per ASTM E1409 (oxygen and nitrogen), E1941 (carbon), or E2371 (plasma emission spectrometry). Mechanical properties - tensile, yield, elongation, reduction of area - are tested per lot.

For larger diameter bar, ultrasonic inspection detects internal defects like voids or inclusions that could become crack initiation sites. Chinese standard GB/T 40324 covers focused ultrasonic testing for large diameter round bars (≥100mm) made from titanium alloys and superalloys.

Grain size and microstructure are checked if specified. Surface condition is visually inspected for cracks, seams, laps, or scale.

 

Documentation

 

Every lot ships with a Mill Test Report (MTR) documenting heat number, chemical analysis, and mechanical test results. The MTR traces back to the original melt.

EN 10204 3.1 is standard. 3.2 with third-party witness is available for OEMs that need independent verification. For ISO 13485 quality systems, the documentation package must link each finished implant to the specific material lot it was machined from.

 

Traceability

 

Implant-grade titanium requires full traceability from melt to finished device. That's not optional for regulatory submissions. FDA recognized ASTM F67 as a consensus standard for surgical implants as early as 1999. The current version, F67-24, remains on the FDA recognition list.

Traceability means each rod carries a heat number that links to the original titanium sponge. Batch records show melting, forging, and drawing history. If a problem surfaces later, the material can be traced and affected devices identified.

 

Customization

 

Standard diameters run 3mm to 50mm. Larger sizes available as forged rod up to 100mm or beyond, depending on mill capability. Lengths are typically 500mm to 3000mm, but cut-to-length is standard - send a cut list and the rod arrives deburred and packaged.

Diameter tolerance follows ISO 286-2. For Swiss machining, h7 is common. For precision surgical instruments, h6. Centerless ground is the default finish.

Anodizing, passivation, and clean-room packaging are available on request.

 

equipment of titanium rod
equipment
 

 

Grade Selection - What Goes Where

 

Grade 1 is the softest and most formable. Used for craniofacial plates, mesh for bone reconstruction, and non-load-bearing dental components. It bends by hand in the OR without cracking.

Grade 2 is the general-purpose medical grade. Used for healing abutments, transfer copings, and components where moderate strength is sufficient. It's also the most commonly specified grade for dental implant fixtures in some markets.

Grade 4 is the strongest pure titanium grade. Used for dental implant fixtures that need osseointegration and some load-bearing capability, and for orthopedic components where Ti-6Al-4V isn't required.

None of the pure titanium grades substitute for Grade 23 ELI in high-stress applications like spinal rods or hip stems. Grade 4 pure titanium has 550 MPa tensile; Grade 23 ELI has 860 MPa. The gap matters when cyclic loading is involved.

 

FAQ

 

 

titanium pure rod

01.What's the difference between ASTM F67 and ASTM F136?

A: ASTM F67 covers unalloyed (pure) titanium - Grade 1 through 4. ASTM F136 covers Ti-6Al-4V ELI alloy (Grade 23). F67 for soft, formable, corrosion-resistant parts. F136 for load-bearing, high-fatigue applications.

02.Can I use ASTM B348 industrial rod for implants?

No. ASTM B348 is commercial grade. ASTM F67 has tighter process control, stricter surface inspection, and full traceability from ore to finished product. Regulators won't accept B348 for implant submissions.

03.What diameter tolerance can you hold?

h7 is standard for most implant machining. h6 is available for precision instruments. Centerless ground rod holds tighter tolerances than turned.

04.What documentation comes with the rod?

MTR with heat number, chemical analysis, mechanical test results, ASTM F67 certificate. EN 10204 3.1 standard. 3.2 with third-party witness available.

05.Lead time?

Stock sizes 7-10 days. Custom sizes 3-5 weeks. Volume orders longer.

 

ASTM F67 is the standard that makes pure titanium implant-grade. Grade 1 for formability. Grade 2 for general medical. Grade 4 for strength. ASTM F67 Pure Titanium Rod Gr1 Gr2 Gr3 Gr4 Surgical Implant delivers the certified material that surgical device manufacturers build on.

 

Contact

Need a quote on ASTM F67 Pure Titanium Rod Gr1 Gr2 Gr3 Gr4 Surgical Implant? Tell us grade, diameter, length, quantity, and surface finish. Include any cert requirements. We'll reply within 24 hours.

Email: kristy@mt-titanium.com
WhatsApp: +86-17391509251

 

Hot Tags: astm f67 pure titanium rod gr1 gr2 gr3 gr4 surgical implant, China astm f67 pure titanium rod gr1 gr2 gr3 gr4 surgical implant manufacturers, suppliers, factory, Biocompatible Titanium Rod for Surgical Implants, High Strength Medical Titanium Rod for Prosthetics, Medical Titanium Alloy Rods for Orthopedic Screws, Ti 6Al 4V ELI Medical Rod for Pediatric Orthopedics, Titanium Alloy Rods Comply With The ASTM F67 Standard, Trauma Surgery Support Titanium Rod Medical Grade

Send Inquiry